Apparent Activity Measurement of ~(125)I Seed Source for Medical Use in Chamber[J]. Journal of Isotopes, 2004, 17(1): 43-43.
Citation: Apparent Activity Measurement of ~(125)I Seed Source for Medical Use in Chamber[J]. Journal of Isotopes, 2004, 17(1): 43-43.

Apparent Activity Measurement of ~(125)I Seed Source for Medical Use in Chamber

  • The effect of self-shield of ~(125)I seed source, container and form of container as well geometry seat of seed on the apparent activity of ~(125)I seed are investigated. The results show that seed sources are carried out under the given a quite severe condition, when the apparent activity measurement of low energy ~(125)I. So, there are several proposals for user to measure the apparent activity of ~(125)I seed source:the source must be put the center of the source disk and measured one by one, while ~(125)I are measured.
  • loading

Catalog

    Turn off MathJax
    Article Contents

    /

    DownLoad:  Full-Size Img  PowerPoint
    Return
    Return