HU Pan, HE Ming, ZHAO Qingzhang, PANG Yijun, GONG Jie, BAO Yiwen, YOU Qubo, LI Kangning, SU Shengyong, HU Yueming. Measurement Technology of 129I Based on Compact AMS[J]. Journal of Isotopes, 2020, 33(4): 199-206. DOI: 10.7538/tws.2019.youxian.069
Citation: HU Pan, HE Ming, ZHAO Qingzhang, PANG Yijun, GONG Jie, BAO Yiwen, YOU Qubo, LI Kangning, SU Shengyong, HU Yueming. Measurement Technology of 129I Based on Compact AMS[J]. Journal of Isotopes, 2020, 33(4): 199-206. DOI: 10.7538/tws.2019.youxian.069

Measurement Technology of 129I Based on Compact AMS

  • Based on the accelerator mass spectrometry system (AMS) with a terminal voltage of 0.3 MV which was independently designed and constructed by China Institute of Atomic Energy (CIAE), the 129I measurement method has been developed. The high efficiency, high precision and high sensitivity measurement method have been developed by systematic studies on ion transport, gas stripping, charge state selection, background elimination, etc. At present, the transmission efficiency is more than 50%, the measurement accuracy reach 0.4%, and the measurement sensitivity of 129I/127I=1×10-14 has been realized by the compact AMS system.