Monte Carlo Simulation of Dosimetric Parameters for Hybrid PdI Source in Brachytherapy
-
Graphical Abstract
-
Abstract
According to dose calculation formula recommended by AAPM TG-43U1,dose rate constant of Model 6711 125I brachytherapy source was calculated by Monte Carlo method. The calculation results were in good agreement with TG-43U1. Then, dose rate constant,radial dose function and anisotropy function of new hybrid PdI source were calculated by Monte Carlo method.Empiric equations were obtained for radial dose function.
-
-