ZHENG Yu-lai, WANG Qiang, ZHANG Wei-zai. Monte Carlo Simulation of Dosimetric Parameters for Hybrid PdI Source in Brachytherapy[J]. Journal of Isotopes, 2013, 26(3): 148-151. DOI: 10.7538/tws.2013.26.03.0148
Citation: ZHENG Yu-lai, WANG Qiang, ZHANG Wei-zai. Monte Carlo Simulation of Dosimetric Parameters for Hybrid PdI Source in Brachytherapy[J]. Journal of Isotopes, 2013, 26(3): 148-151. DOI: 10.7538/tws.2013.26.03.0148

Monte Carlo Simulation of Dosimetric Parameters for Hybrid PdI Source in Brachytherapy

  • According to dose calculation formula recommended by AAPM TG-43U1,dose rate constant of Model 6711 125I brachytherapy source was calculated by Monte Carlo method. The calculation results were in good agreement with TG-43U1. Then, dose rate constant,radial dose function and anisotropy function of new hybrid PdI source were calculated by Monte Carlo method.Empiric equations were obtained for radial dose function.
  • loading

Catalog

    Turn off MathJax
    Article Contents

    /

    DownLoad:  Full-Size Img  PowerPoint
    Return
    Return