ZHANG Wei-hua, WANG Zhi-qiang, LIU Yi-na, XIAO Xue-fu, LI Chun-juan, LUO Hai-long. Measurement and Simulation of TEPC Microdosimetric Spectra in Gamma Radiation Field[J]. Journal of Isotopes, 2013, 26(3): 144-147. DOI: 10.7538/tws.2013.26.03.0144
Citation: ZHANG Wei-hua, WANG Zhi-qiang, LIU Yi-na, XIAO Xue-fu, LI Chun-juan, LUO Hai-long. Measurement and Simulation of TEPC Microdosimetric Spectra in Gamma Radiation Field[J]. Journal of Isotopes, 2013, 26(3): 144-147. DOI: 10.7538/tws.2013.26.03.0144

Measurement and Simulation of TEPC Microdosimetric Spectra in Gamma Radiation Field

  • According to the different characteristics of microdosimetric spectra measured by tissue equivalent proportional counter (TEPC), the neutron dose equivalent and γ dose equivalent could be distinguished in a unknown neutron and γ mixed radiation field. In order to discriminate the γ radiation dose equivalent from the total value,the pure γ microdosimetric spectra was measured in 60Co、137Cs radionuclide radiation field with TEPC. TEPC microdosimetric spectra in a series of monoenergy γ radiation field were simulated by FLUKA code. All the γ radiation microdosimetric spectra, including measured spectrum in 60Co、137Cs radiation field and that of simulation spectrum by FLUKA code, reveal a trait that the linear energy of γ radiation is basically lower than 10 keV/μm. This trait is the very foundation to discriminate the γ radiation from the mixed radiation.
  • loading

Catalog

    Turn off MathJax
    Article Contents

    /

    DownLoad:  Full-Size Img  PowerPoint
    Return
    Return