LIU Jia-ning, LIU Yi-bing, JIA Juan-juan, XU Wen-ge, ZHANG Xue-feng, HAN Shi-quan. Chemiluminescence Immunoassay for Luteinizing Hormone[J]. Journal of Isotopes, 2010, 23(1): 28-33. DOI: 10.7538/tws.2010.23.01.0028
Citation: LIU Jia-ning, LIU Yi-bing, JIA Juan-juan, XU Wen-ge, ZHANG Xue-feng, HAN Shi-quan. Chemiluminescence Immunoassay for Luteinizing Hormone[J]. Journal of Isotopes, 2010, 23(1): 28-33. DOI: 10.7538/tws.2010.23.01.0028

Chemiluminescence Immunoassay for Luteinizing Hormone

  • The chemiluminescence immunoassay(CLIA) for serum luteinizing hormone was established by using the reaction of luminol with hydrogen peroxide. The measurement range of this method was 1.5 to 200 IU/L, the sensitivity was 0.08 IU/L, the recovery rate was 96.3% to 112.1%, the coefficient correlation of dilution was 0.995, and the intra- and inter-assay coefficient of variability were 4.09%-8.36% and 5.14%-10.23%, respectively. Compared with Beckman CLIA system, the coefficient correlation was 0.975.