Quality Evaluation of TSH IRMA Kit[J]. Journal of Isotopes, 2001, 14(Z1): 223-223. DOI: 10.7538/tws.2001.14.Z1.0223
Citation: Quality Evaluation of TSH IRMA Kit[J]. Journal of Isotopes, 2001, 14(Z1): 223-223. DOI: 10.7538/tws.2001.14.Z1.0223

Quality Evaluation of TSH IRMA Kit

  • The quality of TSH (IRMA) kit is estimated by experiment data obtained in 12 months. The essential results contain B min / T ≤0.3%, B max / B min >190, the correlated coefficient of the standard curve≥ 0.993 which have been processed by linear regression of lg D lg( B/B max ),ABCV≤3.4%. Both CV of the curve slope and intercept are less than 5%. All determined values of QC serum are in reference limitation, the CV and bias of inter and intra are less than 10% and 15% respectively. The sensitivity of the method is 0.046 mIU/L, and the stability which is determined by stored the kit at 4℃ is normal, except B min /T is less higher in two months. The samples are determined byTSH RIA and TSH produced by DPC in order to compare with TSH IRMA and the correlated coefficient of the results are 0.949 4 and 0.986 8 respectively.