Apparent Activity Measurement of ~(125)I Seed Source for Medical Use in Chamber

  • 摘要: 研究了125I种子源的自屏蔽、不同容器和容器的形状、源的几何位置等对125I种子源表观活度测量的影响。结果表明,低能125I种子源表观活度的测量只有在严格规定的条件下才能进行较准确的测量。因此为用户测量125I种子源的表观活度提供几点建议:选用材质为聚丙烯的标准尖底放免管作为测量容器;逐个测量;测量时,源必须放在托盘中心。

     

    Abstract: The effect of self-shield of ~(125)I seed source, container and form of container as well geometry seat of seed on the apparent activity of ~(125)I seed are investigated. The results show that seed sources are carried out under the given a quite severe condition, when the apparent activity measurement of low energy ~(125)I. So, there are several proposals for user to measure the apparent activity of ~(125)I seed source:the source must be put the center of the source disk and measured one by one, while ~(125)I are measured.

     

/