Abstract:
11Ccholine is a normal tracer using for PET/CT, identification of
11Ccholine and analysis its radiochemical purity are important to guarantee the validity and safety of tracer. This study aims to establish a new, rapid and effective performance method for identification of
11Ccholine and analysis of the radiochemical purity in the products. The chromatographic method of radio-IC-flow-count were performed using a Metrosep C 6-150/4.0 column with mobile phase of 6 mmol/L HNO
3+5% acetone aqueous solution, flow rate was 0.9 mL/min. The choline chloride and Na
+ retention time were 8.94 min and 3.04 min,
11Ccholine were 9.25 min. The established separating method using IC-flow-count showed good efficiency, The linear range of calibration curves was 10–100 mg·L
–1 for choline chloride, with the recovery rate ranged from 98.9% to 102.1% and the intrabatch and innerbatch variations of < 7.8%. The radio-IC-flow-count method is simple, valid and quick for the identification of
11Ccholine and analysis of radiochemical purity.