分光光度计法测量氨基聚醚的专属性

Specificity of the Spectrophotometry for Detecting Aminopolyether K2.2.2 in 18F-FDG

  • 摘要: 氨基聚醚(K2.2.2)含量是18F-FDG质控中的关键指标,比较了两种常用测量K2.2.2方法的专属性。分别采用分光光度计和半定量TLC 碘显色法测量了14个样品,其中9个阴性样品、2个阳性样品和3个供试品,并与LC-MS/MS测量对比。结果显示:9个阴性样品经分光光度计法测量均为阳性,K2.2.2的测量结果在6.7~470.0 μg/mL;2个阳性样品结果偏高(53,73 μg/mL),3个供试品的K2.2.2含量在14.3~19.2 μg/mL;半定量TLC 碘显色法测量9个阴性样品结果为阴性,2个阳性样品半定量结果与实际一致,3个供试品的K2.2.2含量低于10 μg/mL;LC-MS/MS法测量的14个样品的结果与半定量TLC 碘显色法的结果一致。以上结果表明,半定量TLC 碘显色法测量K2.2.2的专属性较好,适用于测量18F-FDG溶液中K2.2.2含量。

     

    Abstract: The residue of K2.2.2 was one of the key metric in 18F-FDG quality control. The specificity of the spectrophotometry for K2.2.2 was compared with others. Methods: The 14 samples were determined for the K2.2.2 by the spectrophotometry and iodine vapor TLC stain method, respectively. The results were compared with LC-MS/MS. The spectrophotometry showed that the 9 of negative samples were false positive, 2 of positive samples were higher than its actual K2.2.2 contents, 3 of 18F-FDG injections contained 14.3-19.2 μg/mL for K2.2.2. The iodine-vapor TLC stain method showed that 9 of negative samples were negative, 2 of positive samples were the same with its K2.2.2 contents, 3 of 18F-FDG injections were lower than 10 μg/mL with K2.2.2. The iodine vapor TLC stain results were confirmed by the LC-MS/MS. Results: The specificity of spectrophotometry was poor for K2.2.2. The iodine-vapor TLC stain was simple and good specificity method for K2.2.2 in 18F-FDG injections.

     

/