Abstract:
Gamma reference radiation produced by isotope sources
137Cs and
60Co is very necessary for calibration of radiation dosimeter. For the calibration of area monitoring dosimeter, reference radiation must provide the conventionally true value of the ambient dose equivalent
H*(10). According to the recommendation given in the international standard ISO 4037-4, the conventionally true value of
H*(10)could be obtained by two methods, one is spectrum method, the other is
H*(10) standard ionization chamber methed. In order to establish a standard chamber which could meet the requirements of ISO4037-4, a design for
H*(10) standard chamber was established by MCNP4C Monte Carlo simulation program. The results showed that the energy response of the ionization chamber satisfied the requirement of international standard ISO4037-4 from 15 keV to 1500 keV, the conclusion of this study was good to the establishment of
H*(10) secondary standard ionization chamber.