高效液相色谱法测定18F-FDG注射液中氨基聚醚含量

Quantification of Kryptofix2.2.2 in 18F-FDG by HPLC

  • 摘要: 采用高效液相色谱法测定了18F-FDG注射液中氨基聚醚(K2.2.2)的含量,并对流动相、流速、检测波长进行了选择,确定检测条件为:V(50 mmol/L乙酸铵)∶V(乙腈)=1∶1为流动相,流速为0.5 mL/min,检测波长210 nm,进样量为10 μL。K2.2.2在 1~50 mg/L浓度范围内线性良好,线性回归方程:y=265 939.4x+7 490.7,相关系数γ=0.999,回收率为96.0%~100.3%,精密度小于3.5%。本方法回收率较高,精密度较小,样品用量少,可用于18F-FDG注射液中微量K2.2.2的测定。

     

    Abstract: A high-performance liquid chromatograph method was used to quantify Kryptofix2.2.2 (K2.2.2)in the radiopharmaceutical of 18F-FDG(2-Deoxy-2-18F fluoro-D-glucose). A Waters C18 column ( XTerra RP18 150 mm×3.9 mm,5 μm)was used as stationary phase, V(50 mmol/L ammonium acetate)∶V(acetonitrile)=1∶1 was used as mobile phase and the flow was 0.5 mL·min-1, UV detection wavelength was performed at 210 nm. The calibration curve was established over the rang of 1.50 mg/L, the linear correlation coefficient was 0.999, and the recovery was in the rang of 96.0%100.3%. The method was accurate and using less sample,and can be used for the quality control of K2.2.2 in 18F-FDG injection production.

     

/

返回文章
返回