用于PET衰减校正的68Ge线源的制备

Preparation of 68Ge Line Source Used for PET Attenuation Correction

  • 摘要: 以硅胶为载体,采用吸附法制备了一种用于PET衰减校正的68Ge线源,并对制备的68Ge线源进行了质量检验。结果表明,硅胶对68Ge的吸附率高、吸附均匀性好,68Ge线源密封良好,无泄露、无表面污染、活度分布均匀,各项物理参数均达到国外同类产品水平。

     

    Abstract: 68Ge line sources used for PET attenuation correction were introduced by adsorption method with silical gel as substrate, and the quality examination of 68Ge line source was performed. The results showed that the adsorption rate of silical gel for adsorbing Ge was high, and the distribution uniformity of the 68Ge on silical gel was well, the 68Ge line source was sealed well, the surface was pollution-free, and the activity of the 68Ge line source was distributed evenly. The physics properties of the source achieved the same level of the product abroad on the whole.

     

/