傅里叶变换红外光谱法测定D2O浓度
Determination of D2O Concentration by Fourier Infra-red Spectrophotometry
-
摘要: 建立了傅里叶变换红外吸收光谱测定高、中、低浓度重水的方法。该法具有快速、简便、样品无需处理即可直接测定的优点。方法精密度高,D2O含量(摩尔分数)为99.84%时,精密度优于0.0006%;测量范围宽,D2O含量检测范围为99.92%~0.015%;样品用量少,0.2 mL样品即可获得准确结果。Abstract: A method to determine D2O concentration (mole fraction)of high, medium and low heavy water with fourier infra-red spectrophotometer was established. Its outstanding advantages were rapid, convenient, less sample (0.2mL sample), high precision, direct determination without pretreatment. For 99.84 % D2O, the precision was better than 0.0006%. The determined range of D2O was 99.92%-0.015%。