Abstract:
The determination of pertechnetate in techetium99Tc methylenediphosphonate injection by HPLC is proposed.The effects of analytical column, eluent and detection wavelenegth are studied. The detection conditions are on a XTerra RP18 column, using 0.03mol/L HAc-NaAc as mobile phase with a flow rate of 1.0 mL/min and detection wavelenegth of 248nm.The retention time of 99TcO4- is 2.2min and detection limit of 99Tc is 110-3μg.The range of the recovery is 90%~97%,and relative standard diversion is less than 2%.