乏氧显像剂~(99)Tc~m-N4AIPO药盒的研制

Preparation of the Hypoxia-imaging Agent ~(99)Tc~m-N4AIPO Kit

  • 摘要: 系统考察了反应时间、温度、pH、各种成分用量对99Tcm-N4AIPO标记率的影响。在最佳条件下,标记率为95%。同时,利用高效液相色谱法(HPLC)建立药盒中主药(1-氨基-3-(4-硝基咪唑)丙醛肟,N4AIPO)含量的测定方法。观察药盒在不同储藏环境下的稳定性,结果表明药盒在0~4℃冰箱中保存6月以内、室温条件下保存2月以内或于40℃水浴中存放10天以内,可保证放化纯度>90%。说明药盒稳定可靠。另外,异常毒性实验结果表明药盒的毒性低。

     

    Abstract: The effects of reaction time, temperature, pH and amount of the components on the labeling efficiency of ~(99)Tc~m-N4AIPO are systematically investigated. The labeling efficiency is 95% under the optimum labeling conditions. The effective quantity of N4AIPO in kit is determined by HPLC. The stability of the kit under various stock conditions is also examined. The results show that the kit is stable, and the radiochemical purity is keeping more than 90%, or at 0~4 ℃ for 6 months, at room temperature for 2 months, or in 40 ℃ water bath for 10 days. In addition, the acute toxicity of the kit is low.

     

/