影响~(99)Tc~m-MIBI心/肝比值的因素

  • 摘要: 通过使用不同厂家 ̄(99)Tc ̄m-MIBI(下称MIBI)药盒对113例疑有心肌损伤病人进行SPECT断层。采用心/肝放射性比作为心肌对MIBI摄取的参数,对比分析了MIBI药盒的制备、还原剂、就检顺序与心/肝比的关系。结果表明:心/肝比与MIBI药盒还原剂类型无关,在60-120min内,与就检先后顺序无关,呈个体之间的差异。当MIBI的放射化学纯度>95%时,心/肝比与之不相关。

     

    Abstract: ANALYSIS OF THE INFLUENCE FACTORS ON MYOCARDIUM TO LIVER RATIO OF ~(99)Tc~m-MIBI Zhang Jinming Tian Jiahe Zhao Xiaoguang Liu Zhilai(Chinese PLA Generrol Hospitals,Beijing 100853)Wang Shengcai(Jinhuangong Hospital, Datong Coal-Mine of Shaniprovince,036016)ABSTRACT The resting myocardicaL SPECT by different MIBI kits for 113 cases of normal and ischemic heartdisease is performed. The preparation of kits, reducting agents,and the time of imaging are checkedusing myocardium to liver(M/L)ratio as the parameter of myocardium uptake of MIBI.It showsthat MIBI uptake of myocardium has no relation with the reducing agents,the time of imaging between 60-120 min,while M/L ratio varied in every individual case. The ratio of M/L has also norelation to the radiochemical Purity(RCP)as long as RCP is more than 95%.Key words MIBI myocardium to liver ratio radiochemical purity

     

/