Abstract:
MEASUREMENT OF ABSORBED DOSE ON RADIATION STERILIZATION OF MEDICAL ARTICLESLuo Riyou;Wen Zhutang(chinese Nationat Analytical centre,Guangzhou 510070)ABSTRACT In comparison to the results measured by FrjCke dosimeter of natonal standard(called Fricke me-ter in the following text),Alanine Dosimeter of National Dose Assurance Service(called NDAS in thefollowing text)and our dosimeter the errors were±2.0%and±0.5/respectively,Determined D_(10)values of indication bacillus E_(601)(bacillus pumilus)and HBSAg(surface antigen)were 1.69kGy and47.1 kGy respectively.The dose homogeneous degree was 1.33─1.84.Key words E_(601)(bacillus pumilus)HBsAg(surface antigen)D_(10)value of indication bacillusdose homogeneous degree